Value mode (for injection pump M200 only); block pressure alarm limit: 50mmHg~1000mmHg. Alarms: alarm for residual indicator, alarm for infusion completion, alarm for tube block,
alarm for Unclamped barrel, alarm for improperly installed plunger, alarm for speed out of range, alarm for output volume equivalent to limited amount, alarm for power cord falloff,
alarm for battery used Up, alarm for system error, alarm for neglected operation
Power supply
AC power supply: 100V-240V~,50/60Hz
M200A built-in battery: 1V rechargeable lithium-ion battery; the fully charged battery allows over 8.0h working time at the rate of 5ml/h.
M200 base built-in battery: DC11.1V rechargeable lithium-ion battery; the fully charged battery allows the base to work for over 8h .
Operating environment conditions:
Range of ambient temperature: 5℃~40℃
Range of relative humidity: 20%~90%
Range of atmospheric pressure: 700hPa~1060hPa
Power supply: 100V-240V~,50/60Hz
Built-in battery: DC 1V
The syringe pump should work in an environment without strong impact,
vibration or corrosive gas and where water and other fluids cannot flow into the device.
Transportation and storage conditions:
Range of ambient temperature:﹣40℃~+60℃
Range of relative humidity: ≤90%
Range of atmospheric pressure: 700hPa~1060hPa