Description
Feature:
- Standard Parameter : ECG, NIBP, SpO2, Temp, Resp.
- Fully portability with steam line handle
- High brightness 12.1” TFT color LCD
- Big figure display
- ECG waveform of 7-lead display in one screen
- Compatible and self-identify 3/5-lead ECG cable
- NIBP dual over pressure protection
• Isolated floating, anti-defibrillation protected and
anti-high frequency interference
- 72 hours review trend
- 480 hours review trend with external memory
- Optional parameters & Configuration :
2-IBP, EtCO2, 2-TEMP, Recorder, 12-lead ECG,
Central Monitoring network S/W, Wireless Network,
VGA output, Touch LCD - Built-in rechargeable battery
- Dimension (mm) : 310(L) x 140 (W) x 263 (H)
- Weight : 3.8 Kg
Technical Specifications
| Classfication | |
| Anti-electrochock type | Class lequipment and internal powered equipment |
| EMC type | Class A |
| Anti-electroshock degree | ECG(RESP), SPO2, NIBP, IBP, TEMP, CO2, CF |
| Specifications | |||||
| Size Monitor | 310 X 140 X 263 mm | ||||
| Weight Monitor | 3.8kg | ||||
| Temperature | Working | 5 ~ 40℃ | |||
| Transport / Storage | -40 ~ +55℃ | ||||
| Humidity | Working | 30% ~ 75% | |||
| Transport / Storage | ≤ 95% (no coagulate) | ||||
| Altitude | Working | 700hPa ~ 1060hPa | |||
| Transport / Storage | -500hPa ~ 1060hPa | ||||
| Power Supply | 100 ~ 240 VAC, 50/60Hz, Pmax = 150VA, FUSE T1.6A | ||||
| Display | Device | 12.1 in. color TFT, 3 LED / 800 X 600 dots | |||
| Touch LCD(Option) | viewing angle ( up : 70º, down : 70º, Left : 50º, Right : 60º and the best angle is 60º) | ||||
| Messages | – 8 Wave form Maximum
– 1 Power LED (Green) – 3 Soung Mode corresponding Alarm Mode |
– 1 Alarm LED (Yellow / Red)
– 1 Battery Charge LED (Yellow) |
|||
| Signal Interface | ECG Output | BNC | |||
| Amplitude | 1 mV | ||||
| Accuracy | < 5% | ||||
| Impedance | 50Ω | ||||
| Signal Delay | < 20ms | ||||
| Battery | – Rechargeble 3.7 A / Hr 7.4 V Li battery
– Operating time under the normal use and full charge greater than 120 min. – Operating time after the first alarm of low battery will be about 5 min. |
||||
| Recall | Trend Recall | – Short : 1hrs, 1 second Resolution
– Long : 72hrs, 1 min Resolution |
|||
| Alarm Event Recall | 71 alarm events of all parameters and 8/16/32 seconds of corresponding waveform. | ||||
| NIBP Measurement
Recall |
At least 400 NIBP measurement data | ||||
| SD card | – 72 hrs ECG waveform | – 480 hrs TREND review | |||
| ECG | Lead mode | 5 Leads (R, L, F, N, C or RA, LA, LL, RL, V) | |||
| Lead selection | Ⅰ, Ⅱ, Ⅲ, avR, avL, avF, V | ||||
| Waveform | 2CH | ||||
| Lead mode | 3 Leads (R, L, F or RA, LA, LL) | ||||
| Lead selection | Ⅰ, Ⅱ, Ⅲ | ||||
| Waveform | 1ch | ||||
| Gain | x 2.5mm / mV, x5.0mm/mV, x10mm/mV, x20mm/mV | ||||
| HR | – Measuring Range : 15 ~ 300 bpm
– Alarm Range(PED/NEO) : 15~350bpm – Resolution : 1bpm |
– Alarm Rangle(ADU) : 15~300bpm
– Accuracy : ±1% or ±1bpm, which great – Sensitivity : >200uV P-P |
|||
| Differental input
Impedance |
>5MΩ | ||||
| CMRR | – Monitor : ≥100dB
– Operation : ≥100dB – Diagnosis : ≥60dB |
||||
| Electode offset
potential |
±300mV | ||||
| Leakage Current | <10uA | ||||
| Baseline Recovery | <5s After Defi | ||||
| ECG Signal Range | ±8mV(Vp-p) | ||||
| Bandwidth | – Surgery : 1 ~ 20Hz (+0.4dB, -3dB)
– Monitor : 0.5Hz ~ 40Hz (+0.4dB, -3dB) – Diagnostic : 0.05Hz~75Hz (+0.4dB, -3dB), 76Hz~150Hz(+0.4dB, -4.5dB) |
||||
| Calibration Signal | 1mV(Vp-p), ±5% Accuracy | ||||
| ST Segment
Monitoring Range |
Measure Range : -0.6 ~ +0.8mV | ||||
| Alarm Range : -2.0 ~ + 2.0mV | |||||
| ARR Detecting | Type : ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC,
TACHY, BRADY, MISSED BEATS, PNP, PNC |
||||
| Alarm : Available | |||||
| Review : Available | |||||
| RESPARATION | Method | Impedance between R-F(RA-LL) | |||
| Differential Input
Impedance |
>2.5MΩ | ||||
| Measuring
Impedance Range |
0.3 ~ 5.0Ω | ||||
| Base line Impedance
Range |
0.1 ~ 2.5KΩ | ||||
| Bandwidth | 0.3 ~ 2.5Hz | ||||
| Resp. Rate | – Measuring Range : 0 ~ 120rpm
– Alarm Range(PED/NEO) : 0 ~ 150 rpm – Accuracy : ±2rpm |
– Alarm Page(ADU) : 0 ~ 120rpm
– Resolution : 1rpm |
|||
| Apnea Alarm | 10 ~ 40S | ||||
| NIEP | Method | Oscillometric | |||
| Mode | Manual, Auto, STAT | ||||
| Measuring
Interval in AUTO Mode |
1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240, 480, 960 min | ||||
| Measuring
period in STAT Mode |
5min | ||||
| Alarm Type | SYS, DIA, MEAN | ||||
| Measuring and alarm Range | |||||
| Adult mode | – SYS : 40 ~ 270mmHg
– DIA : 10 ~ 215mmHg – MEAN : 20 ~ 235mmHg |
||||
| Pediatric mode | – SYS : 40 ~ 200mmHg
– DIA : 10 ~ 150mmHg – MEAN : 20 ~ 165mmHg |
||||
| Neonatal mode | – SYS : 40 ~ 135mmHg
– DIA : 10 ~ 100mmHg – MEAN : 20 ~ 110mmHg |
||||
| Resolution | Pressure : 1mmHg | ||||
| Accuracy | – Pressure(Maximum Mean error) : ±5mmHg
– Pressure(Maximum Standard deviation) : ±8mmHg |
||||
| Overpressure
protection |
– Adult mode : 297±3mmHg
– Pediatric mode : 240±3mmHg – Neonatal mode : 147±3mmHg |
||||
| SpO2 | Measuring
Range |
0 ~ 100% | |||
| Alarm Range | 0 ~ 100% | ||||
| Resolution | 1% | ||||
| Accuracy | – 70 ~ 100% ±2%
– 0 ~ 69% unspecified |
||||
| Actualization interval | about 1sec | ||||
| Alarm Delay | 10sec | ||||
| Pulse Rate | – Measuring and Alarm Range : 0 ~ 250bpm
– Resolution : 1bpm – Accuracy : ±2bpm |
||||
| TEMPERATURE | Channel | 2 | |||
| Measuring and
Alarm Range |
0 ~ 50℃ | ||||
| Resolution | 0.1℃ | ||||
| Accuracy | ±0.2℃ | ||||
| Actualization
interval |
about 1sec | ||||
| Average Time
Constant |
<10sec | ||||
| IBP (Option) | Channel | 2 | |||
| Label | ART, PA, CVP, RAP, LAP, ICP, P1, P2 | ||||
| Measuring
range |
-10 ~ +300mmHg | ||||
| Alarm range | -50 ~ +350mmHg | ||||
| Press Sensor | – Sensitivity : 5uV/V/mmHg
– Impedance : 300 ~ 3000Ω |
||||
| Resolution | 1mmHg | ||||
| Accracy | ±2% or 1mmHg which great | ||||
| Actualization
Interval |
about 1sec | ||||
| CO2
(Option) |
Method | Infra-red Absorption Technique | |||
| Measuring
mode |
Mainstream and Sidestream | ||||
| Side-stream
mode sampling gas flow rate |
50ml/Min. ± 10ml/Min. | ||||
| Measuring range | CO2 : 0 ~ 150 mmHg
INSCO2 : 0 ~ 150 mmHg AwRR : 2 ~ 150 rpm |
||||
| Resolution | CO2 : 0.1mmHg (0 ~ 69mmHg), 0.25mmHg (70 ~ 150mmHg)
INSCO2 : 0.1mmHg (0 ~ 69mmHg), 0.25mmHg (70 ~ 150mmHg) |
||||
| Accuracy | – CO2 : ±2mmHg, 0 ~ 40mmHg
±5% of reading, 41 ~ 70mmHg ±8% of reading, 71 ~ 100mmHg ±10% of reading, 101 ~ 150mmHg – AwRR : ±1rpm |
||||
| Pace maker (Option) | Availalbe | ||||
| Initialization
Time |
– Mainstrea : Capnogram displayed in less than 15sec at an ambient Temp. of 25 ℃,
full sepcification within 2 min – Sidestrea : Capnogram displayed in less than 20sec at an ambient Temp. of 25℃, full sepcification within 2 min – Mainstream Rise Time : Less then 60ms-Adult Reusable or Single-Patient-Use Airway Adapter Less then 60ms-Infant Reusable or Single-Patient-Use Airway Adapter |
||||
| Acutualization interval | about 1sec | ||||
| Sidestream
Delay Time<</table> |
|||||







Reviews
There are no reviews yet.