Description
Feature:
- Color TFT folding color LCD : 10.2” color TFT
- folding LCD
- 9 crystal pulse wave Doppler transducer
- 152mm wide paper (US/International standard)
- Built-in rechargeable battery
- Accurate fetal movement detection and accounting
- Central monitoring interface
- Type8010 Options: Maternal SpO2, MHR, NIBP,
Twin FHR,ECG, RR, Temp., FAS(Fetal Acoustic Simulator)
Technical Specifications
| Fetal Heart Rate | |
| Transducer | Multi-crystals, pulsed doppler , high sensitively |
| Strength | <5mW/㎠ |
| Working frequency | 1.0MHz |
| Signal processing | special DSP system and modern reorganization. |
| Measurement range | 50~210bpm |
| Alarm Range | |
| High limit | 160,170,180,190bpm |
| Low limit | 90,100,110,120bpm |
| Maximum audio output power | 1.5 Watt |
| Toco | |
| Measurement range | 0~100 units |
| Display | |
| The LCD displays FHR, TOCO, FM, maternal parameters, time, date, and volume, etc., It can store and playback the data. | |
| Power | Type8007 – AC 100-240V, 50/60Hz, 8.4V Li-ION battery Type8010 – AC 100-240V, 50/60Hz, 16.8V Lithum battery |
| Power Consumption | < 20W |
| Dimension | Type8007 : 295W×240D×73H(mm) Type8010 : 320W×350D×85H(mm) |
| Net weight | Type8007: 1.75kg, Type8010: 3.5kg |
| Environment(Working) | Temperature: +5℃ ~ +40℃ |
| Humidity: < 85% | |
| Atmospheric pressure: 86kPa ~ 106kPa | |
| Environment(Transport/storage) | Temperature: -10℃ ~ +55℃ |
| Humidity: < 93% | |
| Atmospheric pressure: 86kPa ~ 106kPa | |
| Transducer Acoustic Output | |
| Under the requirements laid down in IEC 1157, 1992, the peak negative acoustic pressure | |
| does not exceed 1Mpa. The output beam intensity does not exceed 20mW/㎠ | |
| and the spatial-peak temporal-average intensity does not exceed 100mW/㎠ | |
| Maternal SpO2 measure(option) | |
| Measure scope | 70% ~ 99% |
| Measure accuracy | ±3% |
| Pregnant HR Measure(option) | |
| Measure scope | 30bpm~240bpm |
| Measure accuracy | ±2bpm |
| NIBP measure(option) | a) NIBP-SYS (6.7~32.0Kpa)(50 ~ 240mmHg) |
| b) NIBP-DIA (3.4~26.6Kpa)(25 ~ 200mmHg) | |
| c) NIBP-MEAN(2.0~24.0ka(15 ~ 180mmHg) | |
| NIBP Accuracy | ±1.1kPa(±8mmHg)or ±5% of the results, take the bigger one. |
| Maximum Mean error | ±5mmHg |
| Maximum Standard deviation | 8mmHg |
| Pulse Rate Accuracy | Maximum Mean error : ±2bpm |
| Measure accuracy | ±2bpm or ±5% of the results, take the bigger one. |
| Measure mode | Manual / Auto |
| ECG(option) | |
| CMRR: | ≥90dB |
| Heart Rate Measurement and Alarm Range | 15-300bpm |
| Accuracy | ±1bpm |
| Lead Selection | I,II,III,aVR,aVL,aVF |
| Input | 3 / 5-lead ECG cable |
| Sweep Speed | 6.25, 12.5, 25 mm/s |
| Amplitude Selection | x1/4, x1/2, x1, x2, x4 |
| Frequency Response: | |
| Monitoring | 0.5 to 35Hz (+0.4dB, 3.0dB) |
| Defibrillation Protection | 5kV test |
| Recovery Time Following Defibrillation | Less than 5 seconds |
| Temperature(option) | |
| Measurement and Alarm Limit | 0-50°C |
| Probe | Skin surface |
| Unit | Celsius/Fahrenheit |
| Accuracy | ±0.1°C(25°C ~45°C), ±0.2°C(0°C ~25°C or 45°C ~50°C) |
| Resolution | 0.1℃ |
| Respiration(option) | |
| Measurement Mode | Thoracic Impedance |
| Respiration Rate | 7-150 bpm |
| Waveform Display Speed | 6.25, 12.5, 25mm/s |