The infusion pump should work in an environment without strong impact and vibration, free of corrosive gases, and where water and other fluids cannot intrude into the device.
20 drops/ml infusion set: 0.1-1500ml/h; step increment by 0.1ml/h within the range of 0.1-100ml/h, step increment by 1ml/h when above 100ml/h;
60 drops/ml infusion set: 0.1-300ml/h; step increment by 0.1ml/h within the range of 0.1-100ml/h, step increment by 1ml/h when above 100ml/h;
“d/min” is selected as the unit:
20 drops/ml infusion set: 1-400d/min; step increment by 1 d/min; 60 drops/ml infusion set: 1-300d/min; step increment by 1 d/min;
KVO Rate: 0.1-5ml/h; step increment by 0.1ml/h. Factory default: 5ml/h
Ranges of total volume and Volume LimitTotal volume: 0.1-9999ml; step increment by 0.1ml within the range of 0.1-1000ml, step increment by 1ml when above 1000ml.
Volume Limit: 0.1-9999ml; step increment by 0.1ml within the range of 0.1-100ml, step increment by 1ml when above 100ml.
The thresholds of pressure levels are as follows:High: 900mmHg±100mmHg (120±13.3KPa);